Export references
Non-enhanced Low-tube-voltage High-pitch Dual-source Computed Tomography with Sinogram Affirmed Iterative Reconstruction Algorithm of the Abdomen and Pelvis
Affiliations:
Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences & Peking Union Medical College,Beijing,China,100730
Author bio:
Funds:
DOI:
CLC:
Published Online:28 November 2014,
Published:2014
Accepted:
Scan QR Code
Hao Sun, Hua-dan Xue, Zheng-yu Jin, et al. Non-enhanced Low-tube-voltage High-pitch Dual-source Computed Tomography with Sinogram Affirmed Iterative Reconstruction Algorithm of the Abdomen and Pelvis[J]. Chinese medical sciences journal, 2014, (4): 214-220.
DOI:
Hao Sun, Hua-dan Xue, Zheng-yu Jin, et al. Non-enhanced Low-tube-voltage High-pitch Dual-source Computed Tomography with Sinogram Affirmed Iterative Reconstruction Algorithm of the Abdomen and Pelvis[J]. Chinese medical sciences journal, 2014, (4): 214-220.DOI:
Non-enhanced Low-tube-voltage High-pitch Dual-source Computed Tomography with Sinogram Affirmed Iterative Reconstruction Algorithm of the Abdomen and Pelvis
radiation dose and diagnostic value of the low-tube-voltage high-pitch dual-source computed tomography (DSCT) with sinogram affirmed iterative reconstruction (SAFIRE) for non-enhanced abdominal and pelvic scans. Methods This institutional review board-approved prospective study included 64 patients who gave written informed consent for additional abdominal and pelvic scan with DSCT in the period from November to December 2012. The patients underwent standard non-enhanced CT scans (protocol 1) [tube voltage of 120 kVp/pitch of 0.9/filtered back-projection (FBP) reconstruction] followed by high-pitch non-enhanced CT scans (protocol 2) (100 kVp/3.0/SAFIRE). The total scan time
mean CT number
signal-to-noise ratio (SNR)
image quality
lesion detectability and radiation dose were compared between the two protocols. Results The total scan time of protocol 2 was significantly shorter than that of protocol 1 (1.4±0.1 seconds vs. 7.6±0.6 seconds
P<0.001). There was no significant difference between protocol 1 and protocol 2 in mean CT number of all organs (liver
55.4±6.3 HU vs. 56.1±6.8 HU
P=0.214; pancreas
43.6±5.9 HU vs. 43.7±5.8 HU
P=0.785; spleen
47.9±3.9 HU vs. 49.4±4.3 HU
P=0.128; kidney
32.2±2.3 HU vs. 33.1±2.3 HU
P=0.367; abdominal aorta
44.8±5.6 HU vs. 45.0±5.5 HU
P=0.499; psoas muscle
50.7±4.1 HU vs. 50.3±4.5 HU
P=0.279). SNR on images of protocol 2 was higher than that of protocol 1 (liver
5.0±1.2 vs. 4.5±1.1
P<0.001; pancreas
4.0±1.0 vs. 3.6±0.8
P<0.001; spleen
4.7±1.0 vs. 4.1±0.9
P<0.001; kidney
3.1±0.6 vs. 2.8±0.6
P<0.001; abdominal aorta
4.1±1.0 vs. 3.8±1.0
P<0.001; psoas muscle
4.5±1.1 vs. 4.3±1.2
P=0.012). The overall image noise of protocol 2 was lower than that of protocol 1 (9.8±3.1 HU vs. 11.1±3.0 HU
P<0.001). Image quality of protocol 2 was good but lower than that of protocol 1 (4.1±0.7 vs. 4.6±0.5
P<0.001). Protocol 2 perceived 229 of 234 lesions (97.9%) that were detected in protocol 1 in the abdomen and pelvis. Radiation dose of protocol 2 was lower than that of protocol 1 (4.4±0.4 mSv vs. 7.3±2.4 mSv
P<0.001) and the mean dose reduction was 41.4%. Conclusion The high-pitch DSCT with SAFIRE can shorten scan time and reduce radiation dose while preserving image quality in non-enhanced abdominal and pelvic scans.
Abstract
Objective To investigate the image quality
radiation dose and diagnostic value of the low-tube-voltage high-pitch dual-source computed tomography (DSCT) with sinogram affirmed iterative reconstruction (SAFIRE) for non-enhanced abdominal and pelvic scans. Methods This institutional review board-approved prospective study included 64 patients who gave written informed consent for additional abdominal and pelvic scan with DSCT in the period from November to December 2012. The patients underwent standard non-enhanced CT scans (protocol 1) [tube voltage of 120 kVp/pitch of 0.9/filtered back-projection (FBP) reconstruction] followed by high-pitch non-enhanced CT scans (protocol 2) (100 kVp/3.0/SAFIRE). The total scan time
mean CT number
signal-to-noise ratio (SNR)
image quality
lesion detectability and radiation dose were compared between the two protocols. Results The total scan time of protocol 2 was significantly shorter than that of protocol 1 (1.4±0.1 seconds vs. 7.6±0.6 seconds
P<0.001). There was no significant difference between protocol 1 and protocol 2 in mean CT number of all organs (liver
55.4±6.3 HU vs. 56.1±6.8 HU
P=0.214; pancreas
43.6±5.9 HU vs. 43.7±5.8 HU
P=0.785; spleen
47.9±3.9 HU vs. 49.4±4.3 HU
P=0.128; kidney
32.2±2.3 HU vs. 33.1±2.3 HU
P=0.367; abdominal aorta
44.8±5.6 HU vs. 45.0±5.5 HU
P=0.499; psoas muscle
50.7±4.1 HU vs. 50.3±4.5 HU
P=0.279). SNR on images of protocol 2 was higher than that of protocol 1 (liver
5.0±1.2 vs. 4.5±1.1
P<0.001; pancreas
4.0±1.0 vs. 3.6±0.8
P<0.001; spleen
4.7±1.0 vs. 4.1±0.9
P<0.001; kidney
3.1±0.6 vs. 2.8±0.6
P<0.001; abdominal aorta
4.1±1.0 vs. 3.8±1.0
P<0.001; psoas muscle
4.5±1.1 vs. 4.3±1.2
P=0.012). The overall image noise of protocol 2 was lower than that of protocol 1 (9.8±3.1 HU vs. 11.1±3.0 HU
P<0.001). Image quality of protocol 2 was good but lower than that of protocol 1 (4.1±0.7 vs. 4.6±0.5
P<0.001). Protocol 2 perceived 229 of 234 lesions (97.9%) that were detected in protocol 1 in the abdomen and pelvis. Radiation dose of protocol 2 was lower than that of protocol 1 (4.4±0.4 mSv vs. 7.3±2.4 mSv
P<0.001) and the mean dose reduction was 41.4%. Conclusion The high-pitch DSCT with SAFIRE can shorten scan time and reduce radiation dose while preserving image quality in non-enhanced abdominal and pelvic scans.
Intravenous Contrast Material Administration at High-pitch Dual-source CT Coronary Angiography: Bolus-tracking Technique with Shortened Time of Respiratory Instruction Versus Test Bolus Technique
Effects of Saline Administration, Abdominal Compression, and Prolongation of Acquisition Delay on Image Quality Improvement of CT Urography
Related Author
Kai Sun
Guo-rong Liu
Yue-chun Li
Rui-juan Han
Li-fang Cui
Li-jun Ma
Li-gang Li
Chang-yong Li
Related Institution
Department of Radiology
Department of Cardiology, Baotou Central Hospital
CT BM Clinic Marketing, Siemens Healthcare
Department of Physiology, School of Basic Medical Sciences, Wuhan University