Kai Sun, Rui-juan Han, Li-fang Cui, et al. Feasibility and Diagnostic Accuracy for Assessment of Coronary Artery Stenosis of Prospectively Electrocardiogram-gated High-pitch Spiral Acquisition Mode Dual-source CT Coronary Angiography in Patients with Relatively Higher Heart Rates: in Comparison with Catheter Coronary Angiography△[J]. Null, 2012, (4): 213-219.
Kai Sun, Rui-juan Han, Li-fang Cui, et al. Feasibility and Diagnostic Accuracy for Assessment of Coronary Artery Stenosis of Prospectively Electrocardiogram-gated High-pitch Spiral Acquisition Mode Dual-source CT Coronary Angiography in Patients with Relatively Higher Heart Rates: in Comparison with Catheter Coronary Angiography△[J]. Null, 2012, (4): 213-219. DOI:
Objective To prospectively investigate the diagnostic accuracy for coronary artery stenosis of prospectively electrocardiogram-triggered spiral acquisition mode (high pitch mode) dual-source computed tomography coronary angiography (CTCA) in patients with relatively higher heart rates (HR) compared with catheter coronary angiography (CCA). Methods Forty-seven consecutive patients with relatively higher HR (>65 and <100 bpm) (20 male
27 female; age 55±10 years) who both underwent dual-source CTCA and CCA were prospectively included in this study. All patients were performed CTCA using high pitch mode setting at 20%-30% of the R-R interval for the image acquisition. All coronary segments were evaluated by two blinded and independent observers with regard to image quality on a three-point scale (1: excellent to 3: non-diagnostic) and for the presence of significant coronary stenoses (defined as diameter narrowing exceeding 50%). Considered CCA as the standard of reference
the sensitivity
specificity
positive predictive value and negative predictive value were calculated. Radiation dose values were calculated using the dose-length product. Results Image quality was rated as being score 1 in 92.4% of segments
score 2 in 6.1% of segmentsand score 3 in 1.5% of segments. The average image quality score per segment was 1.064±0.306. The HR variability of patients with image score 1
2 and 3 were 2.29±1.06 bpm
5.17±1.37 bpm
8.88±1.53 bpm
respectively. The average HR variability of patients with different image scores were significantly different (F=170.402
P=0.001). The sensitivity
specificity
positive and negative predictive values were 92.6%
97.0%
87.6%
98.3%
respectively
per segment and 90.0%
95.2%
85.3%
96.9%
respectively
per vessel and 100%
63.6%
90.0%
100%
respectively
per patient. The effective radiation dose was on average 0.86±0.16 mSv. Conclusion In patients with HR more than 65 bpm and below 100 bpm without cardiac arrhythmia
the prospectively electrocardiogram-gated high-pitch spiral acquisition mode with image acquired timing set at 20%-30% of the R-R interval provides a high diagnostic accuracy for the assessment of coronary stenoses combined with a 1.5% of non-diagnostic coronary segments and a radiation dose below 1 mSv.
Objective To prospectively investigate the diagnostic accuracy for coronary artery stenosis of prospectively electrocardiogram-triggered spiral acquisition mode (high pitch mode) dual-source computed tomography coronary angiography (CTCA) in patients with relatively higher heart rates (HR) compared with catheter coronary angiography (CCA). Methods Forty-seven consecutive patients with relatively higher HR (>65 and <100 bpm) (20 male
27 female; age 55±10 years) who both underwent dual-source CTCA and CCA were prospectively included in this study. All patients were performed CTCA using high pitch mode setting at 20%-30% of the R-R interval for the image acquisition. All coronary segments were evaluated by two blinded and independent observers with regard to image quality on a three-point scale (1: excellent to 3: non-diagnostic) and for the presence of significant coronary stenoses (defined as diameter narrowing exceeding 50%). Considered CCA as the standard of reference
the sensitivity
specificity
positive predictive value and negative predictive value were calculated. Radiation dose values were calculated using the dose-length product. Results Image quality was rated as being score 1 in 92.4% of segments
score 2 in 6.1% of segmentsand score 3 in 1.5% of segments. The average image quality score per segment was 1.064±0.306. The HR variability of patients with image score 1
2 and 3 were 2.29±1.06 bpm
5.17±1.37 bpm
8.88±1.53 bpm
respectively. The average HR variability of patients with different image scores were significantly different (F=170.402
P=0.001). The sensitivity
specificity
positive and negative predictive values were 92.6%
97.0%
87.6%
98.3%
respectively
per segment and 90.0%
95.2%
85.3%
96.9%
respectively
per vessel and 100%
63.6%
90.0%
100%
respectively
per patient. The effective radiation dose was on average 0.86±0.16 mSv. Conclusion In patients with HR more than 65 bpm and below 100 bpm without cardiac arrhythmia
the prospectively electrocardiogram-gated high-pitch spiral acquisition mode with image acquired timing set at 20%-30% of the R-R interval provides a high diagnostic accuracy for the assessment of coronary stenoses combined with a 1.5% of non-diagnostic coronary segments and a radiation dose below 1 mSv.