BIOMECHANICAL STUDY ONUNILATERAL SINGLE-PLANE EXTERNAL FIXER
were osteotomized at one-third supra-medium segment. They were fixed by an unilateral adjustable external fixer (UAEF)
to be used as a model of external fixation of tibial fracture (MEFTF). The compression. tension
torsion
antero-posterior and lateral bending strength
and the strength for vertically extracting the pins from the tibia were determined in the MEFTF. Within a certain range of load
the correlation of strain to stress was basically a linear relationship. These data provide a theoretical basis of biomechanics for the improvement of UAEF and for early exercises of fracture patients
such as extend-bending of the joint.raising the limb and walking with a walking stick.