[1].Pitfalls in Positron Emission Tomography/Computed Tomography Imaging:Causes and Their Classifications[J].Chinese Medical Sciences Journal,2009,24(01):12-19.
Tian-ran Li1, Jia-he Tian1, Hui Wang1, et al. Pitfalls in Positron Emission Tomography/Computed Tomography Imaging:Causes and Their Classifications[J]. Chinese medical sciences journal, 2009, 24(1): 12-19.
[1].Pitfalls in Positron Emission Tomography/Computed Tomography Imaging:Causes and Their Classifications[J].Chinese Medical Sciences Journal,2009,24(01):12-19. DOI:
Tian-ran Li1, Jia-he Tian1, Hui Wang1, et al. Pitfalls in Positron Emission Tomography/Computed Tomography Imaging:Causes and Their Classifications[J]. Chinese medical sciences journal, 2009, 24(1): 12-19. DOI:
Objective To describe the pitfalls in positron emission tomography/computed tomography (PET/CT) imaging and classify them according to the principles of their generation. Methods We summarized retrospectively the 18F-fluorodeoxyglucose (FDP) PET/CT imaging pit-falls through reviewing the PET/CT images of 872 patients. The pitfalls were divided into artifacts and infrequent physiological uptake
and the artifacts were further classified according to their causes. Meanwhile
we calculated the incidences of various pitfalls. Whether the PET/CT pitfalls influenced the diagnostic decision was analyzed. The appearances of pitfalls in PET were also described. Results Pitfalls could be found in PET/CT images of 684 (78.4%) patients. Artifacts were found in 664 (76.15%) patients
and could be classified into self-factor artifacts and equipment-or technology-related artifacts. Among self-factor artifacts
respiratory motion (57.5%)
postprandial or hyperglycemia artifacts (2.41%)
and metal or high density matter artifacts (1.38%) were frequent. As for equipment- or technology-related factors
injection point outleakage or radiotracer contamination (13.88%) and truncation artifacts (1.83%) were most common ones. Infrequent physiological FDG uptakes
including fatty uptake
endometrial uptake
and bilateral breast feeding period uptake
were found in 20 (2.29%) patients. Among all pitfalls
the artifacts in 92 (13.4%) patients and infrequent physiological uptakes in 6 (0.88%) patients affected the diagnostic results. Artifact images in PET could be described as hot or cold area and the images of infrequent physiological uptake were always shown as hot area. Conclusions The incidence of pitfall in PET/CT imaging was high and the causes of pitfalls are various. Among all causes that artifacts generated
respiratory motion is the most common. Some pitfalls may disturb clinical physicians’ decision
so it is important to recognize artifacts and physiological uptake
and distinguish them from pathological uptakes.
Objective To describe the pitfalls in positron emission tomography/computed tomography (PET/CT) imaging and classify them according to the principles of their generation. Methods We summarized retrospectively the 18F-fluorodeoxyglucose (FDP) PET/CT imaging pit-falls through reviewing the PET/CT images of 872 patients. The pitfalls were divided into artifacts and infrequent physiological uptake
and the artifacts were further classified according to their causes. Meanwhile
we calculated the incidences of various pitfalls. Whether the PET/CT pitfalls influenced the diagnostic decision was analyzed. The appearances of pitfalls in PET were also described. Results Pitfalls could be found in PET/CT images of 684 (78.4%) patients. Artifacts were found in 664 (76.15%) patients
and could be classified into self-factor artifacts and equipment-or technology-related artifacts. Among self-factor artifacts
respiratory motion (57.5%)
postprandial or hyperglycemia artifacts (2.41%)
and metal or high density matter artifacts (1.38%) were frequent. As for equipment- or technology-related factors
injection point outleakage or radiotracer contamination (13.88%) and truncation artifacts (1.83%) were most common ones. Infrequent physiological FDG uptakes
including fatty uptake
endometrial uptake
and bilateral breast feeding period uptake
were found in 20 (2.29%) patients. Among all pitfalls
the artifacts in 92 (13.4%) patients and infrequent physiological uptakes in 6 (0.88%) patients affected the diagnostic results. Artifact images in PET could be described as hot or cold area and the images of infrequent physiological uptake were always shown as hot area. Conclusions The incidence of pitfall in PET/CT imaging was high and the causes of pitfalls are various. Among all causes that artifacts generated
respiratory motion is the most common. Some pitfalls may disturb clinical physicians’ decision
so it is important to recognize artifacts and physiological uptake
and distinguish them from pathological uptakes.