INFLUENCE OF MIMIC CARDIAC RATE ON HYDRODYNAMICS OF DIFFERENT MECHANICAL PROSTHETIC CARDIAC VALVES IN VITRO
中国医学科学杂志(英文版)2005年0卷第2期 页码:138-141
Updated:2024-04-10
|
Share with wechat friends or circle of friends
Export references
INFLUENCE OF MIMIC CARDIAC RATE ON HYDRODYNAMICS OF DIFFERENT MECHANICAL PROSTHETIC CARDIAC VALVES IN VITRO
Affiliations:
Author bio:
Funds:
DOI:
中图分类号:R654.2
纸质出版日期:2005
Accepted:
Scan QR Code
INFLUENCE OF MIMIC CARDIAC RATE ON HYDRODYNAMICS OF DIFFERENT MECHANICAL PROSTHETIC CARDIAC VALVES IN VITRO[J]. 中国医学科学杂志(英文版), 2005,0(2):138-141.
INFLUENCE OF MIMIC CARDIAC RATE ON HYDRODYNAMICS OF DIFFERENT MECHANICAL PROSTHETIC CARDIAC VALVES IN VITRO[J]. Chinese medical sciences journal, 2005, 0(2): 138-141.
INFLUENCE OF MIMIC CARDIAC RATE ON HYDRODYNAMICS OF DIFFERENT MECHANICAL PROSTHETIC CARDIAC VALVES IN VITRO[J]. 中国医学科学杂志(英文版), 2005,0(2):138-141.DOI:
INFLUENCE OF MIMIC CARDIAC RATE ON HYDRODYNAMICS OF DIFFERENT MECHANICAL PROSTHETIC CARDIAC VALVES IN VITRO[J]. Chinese medical sciences journal, 2005, 0(2): 138-141.DOI:
INFLUENCE OF MIMIC CARDIAC RATE ON HYDRODYNAMICS OF DIFFERENT MECHANICAL PROSTHETIC CARDIAC VALVES IN VITRO
摘要
Abstract
Objective To assess the influence of mimic cardiac rate on hydrodynamics of different mechanical prosthetic cardiac valves. Methods US-made CarboMedics bileaflet valve
China-made Jiuling bileaflet valve and C-L tilting disc valve were tested via a pulsatile flow simulator in the aortic position. Testing conditions were set at mimic cardiac rates of 55 bpm
75 bpm
100 bpm with a constant mimic cardiac output of 4 L/min. The mean pressure differences (ΔP)
leakage volumes (LEV) and closing volumes (CLV) across each valve
and effective orifice areas (EOA) were analyzed. Results Within physiological range
ΔP
LEV
and CLV decreased as mimic cardiac rate increased
with a large extent of variance. EOA increased along with an increase in mimic cardiac rate. It was a different response in terms of cardiac rate alteration for different types of mechanical prosthetic cardiac valves. Conclusion Mimic cardiac rate change affects hydrodynamics of mechanical prosthetic cardiac valves. Within physiological range
the hydrodynamic of prosthetic bileaflet valve is better than that of tilting disc valve.