[1]华伟,Harry Mond.LONG TERM PERFORMANCE OF STEROID-ELUTING LEAD IN DUAL CHAMBER PACING[J].Chinese Medical Sciences Journal,1997(02):118-120.
LONG TERM PERFORMANCE OF STEROID-ELUTING LEAD IN DUAL CHAMBER PACING[J]. Chinese medical sciences journal, 1997, (2): 118-120.
<正> The atrial pacing lead is believed having higher stimulation thresholds and long-term complication rates than ventricular lead
this being one of the factors limiting the use of dual chamber pacing. A prospective study was undertaken to evaluate both atrial and ventricular bipolar lined steroid-eluting leads in long-term dual chamber pacing. There are 81 pairs of leads (Medtronic Capsure SP) used in 81 patients. Bipolar atrial and ventricular stimulation thresholds were measured immediately post implantation and 1
3
6
12
18 and 24 months. All leads demonstrated low mean stimulation thresholds during the follow-up and more than 94 % of leads could be paced chronically in the atrium and ventricle at 2.5 volts. In conclusion
atrial and ventricular steroid-eluting leads gave excellent stimulation thresholds allowing low energy long-term dual chamber pacing.
The atrial pacing lead is believed having higher stimulation thresholds and long-term complication rates than ventricular lead
this being one of the factors limiting the use of dual chamber pacing. A prospective study was undertaken to evaluate both atrial and ventricular bipolar lined steroid-eluting leads in long-term dual chamber pacing. There are 81 pairs of leads (Medtronic Capsure SP) used in 81 patients. Bipolar atrial and ventricular stimulation thresholds were measured immediately post implantation and 1
3
6
12
18 and 24 months. All leads demonstrated low mean stimulation thresholds during the follow-up and more than 94 % of leads could be paced chronically in the atrium and ventricle at 2.5 volts. In conclusion
atrial and ventricular steroid-eluting leads gave excellent stimulation thresholds allowing low energy long-term dual chamber pacing.